NOTE
Procedure
Set the DETECTOR POLARITY switch on the HP 3048A Option K21 to
match the polarity of the detector.
The following procedure assumes familiarity with HP 3048A operation.
Initial configuration of the test system and DUT is done in steps 1 to 5.
Calibration is done in step 6, and an AM noise measurement is made in step
7.
Step 1 - Define the measurement
Specify measurement type and frequency range
Measurement type: AM Noise
Start freq: 1 Hz
Stop freq: 40 MHz
Averages: 4
Enter source and interface parameters
Carrier frequency: 12.6 GHz
Detector/discr input frequency: 12.6 GHz
Phase detector: External Phase/AM Detector
Specify a calibration process
Calibration technique: Derive from Double Sided Spur.
Spur Amplitude: –40 dBc
Spur Frequency: 1 kHz
Specify the control of signal sources
DUT SOURCE: USER'S SRCE, MANUAL CONTROL
AM DETECTOR: USER'S DEV., MANUAL CONTROL
CAL SOURCE: USER'S SRCE, MANUAL CONTROL
Define the graph of results
Title: Test 9 - Tunable Source AM Noise and Spurs at 12.6
GHz
Minimum X coordinate: 1 Hz
Maximum X coordinate: 40 MHz
Test 9 - AM Noise and Spurs (RF)
HP 70427A/HP 70428A User's Guide 14-69
Performance Tests