Characteristics
Operating
considerations
Figure 11-6
RF ANALYZER output
This connector provides a monitoring point for the downconverter's input, IF
output, or for a signal connected to the rear-panel SPECTRUM
ANALYZER input from 11848 connector.
Output impedance: 50 Ω
Signal routing
Press the
softkey, then the
Amplitudes
to monitor is underlined. The choices are
connects the rear-panel SPECTRUM ANALYZER input from 11848
connector to the RF ANALYZER connector.
Terminating
Select the
choice if the connector is not used.
-
Terminate the IF output in 50Ω when monitoring the input. Leaving it
unterminated can cause frequency response ripples in the monitored input.
HP 70427A RF Analyzer
Connectors and Indicators (Front- and Rear-Panel)
softkey until the signal you want
VIEW
,
, and
IF
IN
HP 70427A/HP 70428A User's Guide 11-13
RF ANALYZER output
. The last choice
-