Performance Tests
Test 6 - Phase Noise and Spurs of Downconverter
5. Set the reference source output frequency to 2.4 GHz
a. Press the
Output
b. Press the
softkey.
FREQ
c. Enter 2.4 with the numeric keypad.
d. Press the
softkey.
GHz
Measure the reference source output with a power meter and if necessary,
adjust it to +10 dBm ±0.5 dB. Disconnect the cable at the DUT's input
connector and connect the power meter to the end of the cable.
Adjust the DUT's output power for +10 dBm, ± 0.5 dB.
a. Press the
Output
b. Press the
POWER
c. Adjust the power with the knob (0.1 dB steps) or the ↑ ↓ keys (1 db
steps). Note that when entering 0.1 dB steps the actual power may
change as much as 0.2 dB.
d. If necessary, make the DUT module active by pressing the
key followed by the
display the module's softkeys.
6. Set the DUT input frequency to 2.4 GHz
a. Press the
Freqs
b. Press the
INPUT FREQ
c. Enter 2.4 with the numeric keypad.
d. Press the
softkey.
GHz
7. Run the phase-noise measurement
When the phase-noise measurement program stops to prompt a zero-beat
adjustment, do the following:
a. Make sure the HP 70428A Option K01 is turned ON.
b. Adjust the pot on the HP 70428A Option K01 for a zero beat within
4.8 KHz. If necessary, switch the NORMAL/REVERSE switch and
turn the pot in the other direction.
c. Press
to continue measurement.
Proceed
8. Print a graph and parameter summary
On the HP 3048A controller press the
softkey.