Performance Tests
Equipment Required
Table 14-1 shows the test equipment required and their critical
specifications.
Table 14-1
Equipment Required
Instrument Type
Measuring Receiver and
Sensor Module
Phase Noise Measurement
System with Spectrum
Analyzer
Reference Source
Tune Voltage Supply
Balun
Filter
Spectrum Analyzer
μ Wave AM Source
Pulse Generator
Coaxial Detector
1 dB Step Attenuator
14-2 HP 70427A/HP 70428A User's Guide
Critical Specifications
Frequency Range: 2.4 GHz to
25.8 GHz Power Range: +10
to +16 dBm Accuracy: ± 0.5
dB AM Measurement: Depth
to 40% AM Accuracy: 1% of
reading
The Performance Tests for
SSB Phase Noise and AM
Noise are complex and the
procedure has been written
specifically using the HP
3048A, no substitutions are
recommended.
Frequency Range: 2.4 to 25.8
GHz φ Noise ≤ Device under
test.
These accessories are
especially designed and
shielded for φ Noise and AM
Noise measurements. No
substitutions are
recommended.
Frequency Range:
100 kHz to 22 GHz
Frequency: 18 GHz
AM Depth: 40%
Output Power: –5 dBm
Square Wave to 10 kHz:
Level Adjustable to 1 Vpk.
Frequency Range to 25.8
GHz
0 –12 dB; 1 dB steps
Recommended Model
HP 8902A
HP 11792A
HP 3048A/201/H26
HP 3585B
HP 5316B (optional)
HP 70428A
HP 70428A K01
HP 70427A K02
HP 3048A K21
HP 8562A
HP 8566B
HP 70XXX series
HP 83640A
HP 8116A
HP 33330C
HP 355C

Advertisement