Table 26. Possible Solutions for Unusual Behavior
The touchscreen or the keys do not respond.
Press and hold
until the tester turns off. Then turn on the tester. If the
problem continues, install the latest version of software in the tester.
The tester will not turn on even though the battery is charged.
The safety switch or the fuse in the battery is possibly open. Wait a few
seconds, then try again. If the tester does not turn on, the battery is possibly
bad and must be replaced. The tester can operate on AC power if the
battery does not operate.
The OTDR Port Connection Quality gauge is not in the "Good" range.
 The fiber endface in the OTDR connector is possibly dirty or damaged.
Clean the connector and inspect the endface with a video probe. If an
endface is damaged, contact Fluke Networks for service information.
 Remove the fiber adapter from the module and inspect the adapter for
damage. Make sure that the white plastic ring inside the center tube
shows no damage.
Test results appear to be incorrect.
 The tester possibly has incorrect settings. Make sure you selected the
correct cable type and test limit. If you did a Manual OTDR test, do an
Auto OTDR test to see if you get the correct results.
 The tester is connected to the link with the wrong type of fiber. Make
sure you use the correct type of launch/tail cords.
See Chapter 15 "Diagnose Problems in Fiber Links" for information on how
incorrect settings and connections can possibly affect your test results.See
the Technical Reference Handbook for information on how incorrect
settings and connections can possibly affect your test results.
The message "Selected test limit or fiber type is not valid for this test"
shows.
 The test limit requires two wavelengths, but you selected only one. Select
the both wavelengths on the TEST SETUP screen.
 The installed module does not support the selected fiber type or test
limit. For example, the test limit has a singlemode fiber type, but a
multimode module is installed. Select a different fiber type or test limit
or install a different module.
Before You Send a Main Tester to a Service Center
Chapter 20: Maintenance
599