Set The Reference - Fluke Versiv Series Technical Reference Handbook

Hide thumbs Also See for Versiv Series:
Table of Contents

Advertisement

Versiv Series Cabling Certification Product Family
Technical Reference Handbook

Set the Reference

The reference procedure for twisted pair cable sets the baseline
for insertion loss, ACR-F, and DC resistance measurements.
Set the reference at these times:
When you want to use the tester with a different module. The
tester can save reference values for eight different pairs of
modules.
When you attach Class F/F
optional DSX-PLA011 TERA
Every 30 days, at minimum. To ensure maximum accuracy of
test results, set the reference daily.
It is not necessary to set the reference when you change the link
interface adapters (unless you attach Class F/F
adapters).
To set the reference
Install DSX modules in the tester and the remote.
1
Turn on the tester and the remote a minimum of 5 minutes
2
before you set the reference.
Set the reference only after the testers are at an
ambient temperature between 50 °F and 104 °F
(10 °C and 40 °C).
Use the appropriate adapters or a reference artifact to
3
connect the main and remote testers together as shown in
Figure 27.
On the home screen, tap TOOLS, then tap Set Reference.
4
On the SET REFERENCE screen tap TEST.
5
82
link interface adapters, such as the
A
adapters.
Note
link interface
A