CertiFiber Pro OLTS Module Specifications

CertiFiber Pro Power Meter Specifications

Input connector
Detector type
Wavelengths
Power measurement
range
Display resolution
Display update rate
Power measurement
uncertainty (accuracy)
Measurement linearity
Re-calibration period
 5 %
100 pW at 850 nm
1.
2. Under the following conditions:
Power level 100 μW (-10 dBm), continuous wave (CW) for absolute power at
850 nm and 1310 nm.
Divergent beam, NA = 0.20 for 50/125 μm and NA = 0.14 for 9/125 μm.
Ambient temperature 23° ± 1 °C.
SC/UPC connector with ceramic ferrule.
After a 5 minute warm-up.
Traceable to NIST.
3. -3 dBm to -55 dBm at 850 nm and 1310 nm. Ambient temperature 23 °C ± 1 °C.
After a 5 minute warm-up.
CertiFiber Pro OLTS Module Specifications
Interchangeable connector adapter
(LC standard)
InGaAs
850 nm, 1310 nm, 1550 nm
850 nm: +10 dBm to -65 dBm
All other wavelengths: +10 dBm to -70 dBm
dB, dBm: 0.01
Linear (nW, μW, mW): 0.001 to 9.999, 10.00 to 99.99,
100.0 to 999.9 depending upon value
1 reading per second
 5 %  32 pW
1,2
<  0.1 dB
3
1 year