Table 2. Power Button's LED Indications for Versiv 2 Testers
Red, steady: The tester is on and the battery is charging. The
LED stays red for a few minutes after the battery status icons
show that the battery is fully charged.
Red, flashing: The tester is off and the battery is charging.
Red steady or alternating red/green: The AC adapter is
connected, but the battery is not installed.
Yellow, steady: The tester is on and the AC adapter is
connected, but there is an issue with the battery:
 The battery voltage is very low. The AC adapter will trickle-
charge the battery until the voltage increases, then charge
the battery at the normal rate.
 The battery will not charge because its temperature is outside
the range of 32 °F to 104 °F (0 °C to 40 °C).
Yellow, flashing: Same as above, but the tester is off.
Yellow, flashes 5 times: This occurs when the tester reboots
after a software update, or if tester reboots because it had an
x5
unrecoverable error.

Check the Battery Status

On a main tester
The battery status icon is in the upper-left corner of the screen:
Battery is full. The LED stays red for a few minutes after the
battery status icons show that the battery is fully charged.
Battery is approximately half full.
If the AC adapter is not connected, the red bar shows that
the battery is low. Connect the AC adapter to charge the
battery and make sure the tester continues to operate.
The red bar also shows if the AC adapter is connected, but
the battery is not installed.