Step 3: Turn Off the Auto Save Function

For this tutorial, you will manually save the test results.
On the home screen, tap the Next Fiber A ID/Next Fiber B ID
3-1
panel.
On the CHANGE ID screen, tap the Auto Save control to make
3-2
it show Off.
Stay on this screen to do the next step.
3-3

Step 4: Enter IDs for the Fibers

On the CHANGE ID screen, tap the FIBER A ID: panel.
4-1
Enter an ID for fiber A, then tap DONE twice.
4-2
The CHANGE ID screen shows the ID you entered for fiber A.
For the Fiber B ID, the tester increments the Fiber A ID by one.
Tap DONE to save the IDs. If you see a message about an ID
4-3
that is not in the list, tap YES.

Step 5: Set Up Launch Fiber Compensation

On the home screen, tap SET COMP.
5-1
On the SET LAUNCH METHOD screen, make sure that
5-2
SmartLoop is selected.
Clean and inspect the OTDR port and the connectors on the
5-3
three launch cords.
Make the connections shown in Figure 16.
5-4
Tap SET.
5-5
When the SET LAUNCH COMP screen shows, make sure the
5-6
tester shows the correct distances for the end of the launch
cord and the start of the tail cord. See Figure 17.
Tutorial: Do a Bi-directional SmartLoop Test