Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Page 3
OPERATOR SAFETY SUMMARY WARNING HIGH VOLTAGE is used in the operation of this equipment LETHAL VOLTAGE may be present on the terminals, observe all safety precautions! To prevent electrical shock hazard, the operator should not electrically contact the output HI or sense HI terminals or circuits connected to these terminals. During operation, lethal voltages of up to 1020 V ac or dc may be present on these terminals.
Connect to Line Power ..................8 Select Line Frequency ..................8 Placement ......................10 Airflow Considerations ..................11 Instruction Manuals ................... 11 5502E Getting Started Manual ..............11 5502A Operators Manual ................11 General Specifications ..................12 Detailed Specifications ..................13 DC Voltage ....................13 DC Current ....................
Page 6
5502E Getting Started AC Voltage, Triangle Wave Characteristics (typical) ........24 AC Current (Non-Sine Wave) ............... 24 AC Current (Non-Sine Wave) (cont.) ............25 AC Current, Square Wave Characteristics (typical) ........25 AC Current, Triangle Wave Characteristics (typical) ........25...
Page 7
Title Page Symbols ........................4 Standard Equipment ....................7 Mains-Power Cord Types Available from Fluke Calibration ........ 10 Verification Tests for DC Voltage (Normal) ............26 Verification Tests for DC Current (AUX) ............. 27 Verification Tests for Resistance ................28 Verification Tests for AC Voltage (Normal) ............
Page 9
List of Figures Figure Title Page 5502E Multi-Product Calibrator ................2 RS-232 Remote Connection ................... 6 Access the Fuse and Select Line Voltage ............... 9 Mains-Power Cord Types Available from Fluke Calibration ........ 10 Allowable Duration of Current >11 A ..............14...
The 300 MHz and 600 MHz options are not compatible with the 5502E. The 5502E does not have SCOPE and TRIG OUT terminals. The SCOPE and TRIG OUT buttons are on the front panel of the 5502E, but they are non-functional and beep when pushed.
Page 12
488.1-1987 and 488.2-1987. • A EIA Standard RS-232 serial data interface to print, show, or move internally stored calibration constants, and for remote control of the 5502E. • A pass-through RS-232 serial data interface to send data to the Unit Under Test...
Multi-Product Calibrator Safety Information Safety Information In this manual, A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. Warning To prevent possible electrical shock, fire, or personal injury: •...
WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. Risk of Danger. Important ...
• Brazil: +55-11-3759-7600 • Anywhere in the world: +1-425-446-6110 To see product information and download the latest manual supplements, visit Fluke Calibration’s website at www.flukecal.com. To register your product, visit http://flukecal.com/register-product. Overload Protection The Calibrator supplies reverse-power protection, fast output disconnection, and/or fuse protection on the output terminals for all functions.
5502E Getting Started Remote Operation (RS-232) There are two rear-panel serial data RS-232 ports: SERIAL 1 FROM HOST, and SERIAL 2 TO UUT (see Figure 2). Each port is dedicated to serial-data communications to operate and control the Product during calibration procedures. For complete information on remote operation, see Chapter 5 of the Operators Manual.
Calibration Service Center (see “Contact Fluke Calibration”). A performance test is in the “Maintenance” section of Chapter 7 in the Operators Manual. If you ship the calibrator to Fluke Calibration, use the initial container. If it is not available, you can get a new container from Fluke Calibration with the Calibrator model and serial number.
If it is necessary to use a different type, refer to Table 3 and Figure 4 for a list and illustration of the line-power plug types available from Fluke Calibration. After you make sure that the line voltage selection is set correctly and that the correct fuse for that line voltage is installed, connect the calibrator to a correctly-grounded three-prong outlet.
Page 19
Multi-Product Calibrator Select Line Frequency Line Voltage Indicator Changing Line Fuse Changing Line Voltage gjh004.eps Figure 3. Access the Fuse and Select Line Voltage...
5502E Getting Started Table 3. Mains-Power Cord Types Available from Fluke Calibration Fluke Calibration Option Type Voltage/Current Number North America 120 V/15 A LC-1 North America 240 V/15 A LC-2 Universal Euro 220 V/15 A LC-3 United Kingdom 240 V/13 A...
• 5502E Getting Started (PN 4238530) One of each manual above comes with the instrument. Refer to the Fluke Calibration Catalog or speak to a Fluke Calibration sales representative (see “Contact Fluke Calibration”) if more printed copies are necessary. The manuals are also available on the Fluke Calibration website.
The following tables list the 5502E specifications. All specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5502E has been turned off. (For example, if the 5502E has been turned off for 5 minutes, the warm-up period is 10 minutes.) All specifications apply for the temperature and time period indicated.
Weight (without options) ........22 kg (49 lb) Absolute Uncertainty Definition ......The 5502E specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total specification of the 5502E for the temperature range indicated.
Formula 60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in amperes. For example, 17 A, at 23 °C could be provided for 60-23-17 = 20 minutes each hour. When the 5502E is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle.
Multi-Product Calibrator Detailed Specifications Resistance Absolute Uncertainty, tcal ±5 °C ±(% of output + floor) Resolution Allowable Current Floor (Ω) Time and temp since ohms Range % of output (Ω) zero cal 90 Day 1 Year 12 hrs ±1 °C 7 days ±5 °C 0 to 10.999 Ω...
5502E Getting Started AC Voltage (Sine Wave) Absolute Uncertainty, tcal Max Distortion and ±5 °C ±(% of output + μV) Noise 10 Hz to Range Frequency Resolution Max Burden 5 MHz Bandwidth ±(% of output + 90 Day 1 Year...
60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in amps. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the 5502E is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle.
60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in amps. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20 minutes each hour. When the 5502E is outputting currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle.
Multi-Product Calibrator Detailed Specifications Capacitance Absolute Uncertainty, tcal ±5 °C Allowed Frequency or Charge-Discharge Rate [1] [2] [3] ±(% of output + floor) Range Resolution Min and Max to Typical Max for Typical Max for 90 Day 1 Year Meet <0.5 % Error <1 % Error Specification...
5502E Getting Started Temperature Calibration (Thermocouple) Absolute Uncertainty Absolute Uncertainty Source/Measure tcal ±5 °C ± Source/Measure tcal TC Type Range °C Range °C Type °C ±5 °C ± °C 90 Day 1 Year 90 Day 1 Year 600 to 800 0.42...
The subsequent paragraphs provide additional specifications for the 5502E Calibrator ac voltage and ac current functions. These specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5502E has been turned off. All extended range specifications are based on performing the internal zero-cal function at weekly intervals, or when the ambient temperature changes by more than 5 °C.
5502E Getting Started AC Voltage (Sine Wave) Extended Bandwidth 1-Year Absolute Uncertainty Range Frequency Max Voltage Resolution tcal ±5 °C Normal Channel (Single Output Mode) 1.0 to 33 mV Two digits, e.g., 25 mV ±(5.0 % of output 34 to 330 mV 0.01 to 9.99 Hz...
Multi-Product Calibrator Additional Specifications AC Voltage (Non-Sine Wave) (cont.) 1-Year Absolute Uncertainty, tcal Square Wave Frequency Max Voltage Resolution ±5 °C, ±(% of output + % of range) Range (p-p) Normal Channel (Single Output Mode) 0.01 to 10 Hz 5.0 + 0.5 Two digits on each range 10 to 45 Hz 0.25 + 0.5...
Multi-Product Calibrator Additional Specifications AC Current (Non-Sine Wave) (cont.) 1-Year Absolute Uncertainty tcal ±5 °C Max Current Square Wave Range p-p Frequency ±(% of output + % of range) Resolution 10 to 45 Hz 0.25 + 0.5 0.047 to 0.65999 mA 45 Hz to 1 kHz 0.25 + 0.25 Six digits...
The limits were calculated by adding or subtracting the 90-day specification from the output value. There is no built-in factor for measurement uncertainty. These performance tests replace those found in the 5502A Operators Manual and only apply to the 5502E. Table 4. Verification Tests for DC Voltage (Normal)
5502E Getting Started Table 7. Verification Tests for AC Voltage (Normal) Range Output Frequency Lower Limit Upper Limit 32.999 mV 3.000 mV 45 Hz 2.977 mV 3.022 mV 32.999 mV 3.000 mV 10 kHz 2.977 mV 3.022 mV 32.999 mV 30.000 mV...
Page 41
Multi-Product Calibrator Performance Tests Table 7. Verification Tests for AC Voltage (Normal) (cont.) Range Output Frequency Lower Limit Upper Limit 3.29999 V 3.00000 V 100 kHz 2.99437 V 3.00563V 3.29999 V 3.00000 V 450 kHz 2.98659 V 3.01340 V 3.29999 V 3.29000 V 1 MHz 2.250 V...
5502E Getting Started Table 8. Verification Tests for AC Current Range Output Frequency Lower Limit Upper Limit 329.99 μA 33.00 μA 32.87 μA 33.13 μA 1 kHz 329.99 μA 33.00 μA 32.60 μA 33.40 μA 10 kHz 329.99 μA 33.00 μA 32.20 μA...
Page 43
Multi-Product Calibrator Performance Tests Table 8. Verification Tests for AC Current (cont.) Range Output Frequency Lower Limit Upper Limit 32.999 mA 19.0000 mA 30 kHz 18.935 mA 19.065 mA 32.999 mA 32.9000 mA 10 Hz 32.849 mA 32.951 mA 32.999 mA 32.9000 mA 1 kHz 32.886 mA...
Page 44
5502E Getting Started Table 8. Verification Tests for AC Current (cont.) Range Output Frequency Lower Limit Upper Limit 2.99999 A 2.99000 A 10 kHz 2.92520 A 3.05480 A 20.5000 A 3.3000 A 500 Hz 3.2954 A 3.3046 A 20.5000 A 3.3000 A...
5502E Getting Started Table 9. Verification Tests for Capacitance (cont.) Test Frequency Range Output Lower Limit Upper Limit or Current 540 μA dc 3.299 mF 2.000 mF 1.9902 mF 2.0098 mF 800 μA dc 3.299 mF 3.000 mF 2.9868 mF 3.0132 mF...