Electromagnetic Compatibility (EMC)

Guidance and Manufacturer's Declaration – Electromagnetic Emissions
Emissions Test
RF Emissions
EN 55011
RF Emissions
EN 55011
Harmonic Emissions
EN 61000-3-2
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
2042480-001 C
EMC Compliance: Electromagnetic Compatibility (EMC)
Changes or modifications to this system not expressly approved by GE Medical
Systems can cause EMC issues with this or other equipment. This system is
designed and tested to comply with applicable regulation regarding EMC and
must be installed and put into service according to the EMC information stated
in this appendix.
WARNING
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation.
WARNING
The equipment or system should not be used adjacent to, or
stacked with, other equipment. If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in the configuration in which it is being used.
The Dash 2500 Patient Monitor is intended for use in the electromagnetic
environment specified below. It is the responsibility of the customer or user to
assure that the Dash 2500 Patient Monitor is used in such an environment.
Compliance
Group 1
The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class A
Class A
The equipment is suitable for use in all establishments other than
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
Complies
Dash 2500 Patient Monitor
Electromagnetic Environment – Guidance
E-3