OP Panel A/N Keyboard Assy (FRU No. 207)

Purpose: This is a description on how to remove and replace OP Panel A/N Keyboard Assy.
8-4-8-1
Tools
Common pillips screwdrivers.
8-4-8-2
Needed Manpower
1persons, 5 minutes + travel
8-4-8-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
page 16.
8-4-8-4
Removal Procedure
1.) Unscrew five (5) screws and remove the HUB Assy EMC guard. Refer to Figure 8-30.
2.) Disconnect two(2) connectors (1-2) from the HUB Board. Refer to Figure 8-31.
8 - 30
7
1
2
Figure 8-30 Removing HUB EMC guard
1
Figure 8-31 Disconnecting 2 connectors