Op Panel Hub Board Assy (Fru No. 209); Direction 2380207, Revision 7 Logiq™ 5 Pro Service Manual - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-4-10

OP Panel HUB board Assy (FRU No. 209)

Purpose: This is a description on how to remove and replace HUB board Assy.
8-4-10-1
Tools
Common pillips screwdrivers.
8-4-10-2
Needed Manpower
1persons, 5 minutes + travel
8-4-10-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
Keyboard Assy should be removed before proceed. For more information, refer to section 8-4-1 on
page 16.
8-4-10-4
Removal Procedure
1.) Unscrew five (5) screws and remove the EMC guard from the HUB Board. Refer to Figure 8-36.
2.) Disconnect six (6) connectors (1-6) from the HUB Board Assy. Refer to Figure 8-37.
7
1
2
Figure 8-36 Open EMC guard
5
6
Figure 8-37 Disconnect 6 connectors
Chapter 8 Replacement Procedures
LOGIQ™ 5 PRO S
ERVICE
3
4
5
1
2
3
4
M
ANUAL
8 - 35

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents