Reassembly

Top Cover

Removal

Revision E
Maintenance: Disassembly Guidelines
Reassemble the power supply reversing the steps for removal. Before replacing the
screws, ensure that the ground wire is routed through the notch in the plastic and not
pinched.
NOTE
It is not necessary to remove the Power Supply prior to removing the top cover.
NOTE
A #10 TORX driver is required for disassembly and assembly of the top cover.
1.
Remove the battery.
2.
Turn the unit over so the bottom side is up and remove the TORX screw
through the hole on the right rear corner of the unit. (This screw is only visible
and accessible with the battery removed.)
MAC 5500 resting ECG analysis system
2020299-020
Screws (3)
Ground Wire
Wiring Harness
TORX
screw
41A
40A
5-9