Monitor Cover Set (Fru No. 102) - GE LOGIQ 5 PRO Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2380207, R
IRECTION
EVISION
8-3-3

Monitor Cover Set (FRU No. 102)

Purpose: This is a description on how to remove the Monitor Cover Set.
8-3-3-1
Tools
Common pillips screwdrivers
8-3-3-2
Needed Manpower
1person, 10 minutes + travel
8-3-3-3
Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 4.
8-3-3-4
Removal Procedure
1.) Remove the Tasklamp Assy, For more information, refer to section 8-3-6 on page 14.
2.) Unscrew three screws (1-2,6) to remove the Monitor Cover left.
3.) Unscrew three screws (3-5) to remove the Monitor Cover right.
4.) Unscrew four screws (7-10) to remove the Monitor Cover front.
5.) Monitor Cover Top can be removed without unscrewing.
1
2
8 - 8
7
3
4
6
Figure 8-9 MONITOR COVER SET DISASSEMBLY
Section 8-3 - Monitor
LOGIQ™ 5 PRO S
5
7
8
M
ERVICE
ANUAL
9
10