To Change the Liner and O-Ring on the Split/Splitless Inlet

N O T E
WA RNING
Maintaining Your GC
Verify that a Guard chip or jumper chip is installed. If not install it
before installing the liner.
1
Gather the following:
Replacement O-ring (See
Split/Splitless Inlet"
Replacement liner
Tweezers
Wrench, hex for changing septum
Wrench, capillary inlet (optional)
Lint-free gloves
Launch the GC maintenance wizard: Maintenance > Inlet >
2
Perform Maintenance > Replace Liner and O-Ring> Start
Maintenance. The wizard will walk through the replacement
procedure. These steps are repeated below.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
3
Slide the locking tab forward (counterclockwise). Lift the
septum assembly straight up and away from the inlet to avoid
chipping or breaking the liner.
Maintaining the Split/Splitless Inlet
"Consumables and Parts for the
on page 92.)
3
99