Columns and Traps
Procedure: Installing capillary columns in the purged packed inlet
Procedure: Installing capillary columns in the purged packed inlet
Before installing a column in this inlet, be sure you have a capillary liner and
glass insert installed. Instructions on choosing and installing this hardware are
in the Inlets volume. If your insulation cup is not installed, begin with Step 1.
Otherwise, begin with Step 4.
Materials required
Column nut and ferrule
Column cutter
Typewriter correction fluid
1/4-inch wrench
Metric ruler
Insulation cup
No-hole ferrule to use as a plug when installing the insulation cup
1. Install a plug in the inlet fitting.
106
Inlet fitting
Capillary liner
Column nut with
no-hole ferrule
Insulation cup
Part 1. Capillary columns