11
Maintaining the Split/Splitless Inlet

To Replace the Gold Seal on the Split/Splitless Inlet

WA R N I N G
174
1
Gather the following:
Replacement gold seal (See
the Split/Splitless Inlet"
Replacement washer
1/4-inch wrench (for column)
1/2-inch wrench
Lint-free gloves
Load the maintenance method and wait for the GC to
2
become ready.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Open the Guard Column enclosure door.
3
Remove the inlet guard column from the inlet. Cap the open
4
end of the column to prevent contamination.
Figure 22 Inlet
"Consumables and Parts for
on page 168.)
Inlet
Troubleshooting and Maintenance Guide