Performance Verification

Introduction

Initial Setup

20
5 — Battery Low Indicator
The SRC-MAX functional tester allows qualified technicians to functionally
test Nellcor O
M
technology-based pulse oximeters and OEM O
XI
AX
technology-based monitors. The technician must perform the test setup
procedure before performing tests 1 through 4. The following is a brief
description of each test:
Test Setup — This procedure establishes the baseline for all the other
tests. The Test Setup procedure must be performed before performing
any or all of the SRC-MAX tests.
Test #1: BPM — This procedure simulates an O
attached to a patient indicating 60 bpm and 200 bpm. The test setup
procedure sets up Test 1 for 60 bpm.
Test #2: SpO
— This procedure simulates an O
2
attached to a patient, indicating 75 percent blood oxygen saturation
and 90 percent blood oxygen saturation. The test setup procedure sets
up Test 2 for 75 percent blood oxygen saturation.
Test #3: Modulation — This procedure simulates an O
attached to a patient indicating low and high pulse strength. The test
setup procedure sets up Test 3 for low pulse strength.
Test #4: Light — This procedure simulates an O
attached to a patient indicating low and high light level passing
through the patient at the sensor site. The test setup procedure sets up
Test 4 for low light level.
10 — Infrared LED Drive Indicator
1. Sensor Port
M
XI
AX
M
sensor
XI
AX
M
sensor
XI
AX
M
sensor
XI
AX
M
sensor
XI
AX