Analog Calibration Setup - Nellcor OxiMax N-595 Service Manual

Hide thumbs Also See for OxiMax N-595:
Table of Contents

Advertisement

Data Port Interface Protocol
NCALL
NORM +
NORM -

Analog Calibration Setup

SETUP
NEXT
NEXT
NEXT
ANALOG
0 VOLT
1 VOLT
STEP
110
5. Press the NCALL softkey.
6. Press the NORM+ (+5 to +12 VDC) or NORM- (-5 to -12 VDC) softkey as
required for your nurse call system.
Analog calibration signals are provided to adjust a recorder to the output of the
instrument. Selectable calibration signals are +1.0 VDC, 0.0 VDC, and Step. For
more information on the analog signals see Analog Output on page 123.
1. Turn on the N-595 by pressing the ON/STANDBY button.
2. Press the SETUP softkey.
3. Press the NEXT softkey.
4. Press the NEXT softkey.
5. Press the NEXT softkey.
6. Press the ANALOG softkey.
7. Press the 0 VOLT, 1 VOLT, or STEP softkey as required.

Advertisement

Table of Contents
loading

Table of Contents