Analog Calibration Setup - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

Data Port Interface Protocol
Analog Calibration
Setup
94
5. Press the NCALL softkey.
6. Press the NORM+ (+5 to +12 VDC) or NORM- (-5 to -12 VDC) softkey
as required for your nurse call system.
Analog calibration signals are provided to adjust a recorder to the output of
the instrument. Selectable calibration signals are +1.0 VDC, 0.0 VDC, and
Step. For more information on the analog signals see Analog Output
page 106.
1. Turn on the N-595 by pressing the POWER ON/OFF button.
2. Press the SETUP softkey.
3. Press the NEXT softkey.
4. Press the NEXT softkey.
5. Press the ANALOG softkey.
6. Press the 0 VOLT, 1 VOLT, or STEP softkey as required.
7. Press the BACK softkey.