Using the Data Port

Setting Nurse Call Relays Normally Open/Closed

Calculating the Analog Voltage Output

118
Data port pins 7 and 15 provide a relay that closes when an alarm is
sounding on the monitor. Pins 8 and 15 provide a relay that opens
when an alarm is sounding. Pin 15 is a common lead for both relays.
The relay operates whether the monitor is operating on AC power or
battery.
The N-600x data port provides analog voltage outputs between
pins 6, 13, 14, and ground (pin 10), which can be used to calibrate
instruments such as a chart recorder. The voltage represents a specific
measured parameter's current value. The voltage differential varies
proportionally from 0 to 1 volt as the pin's parameter varies over its
full range of values, as indicated in Table 9.
Table 9: Analog Pinouts
Pin
Parameter
6
%SpO2
13
Pulse Rate
14
Pleth Waveform
For example, as the current value of %SpO
the voltage from pin 6 to ground (pin 10) varies from 0 to 1 volt. A
voltage of 0.94 volts indicates a current %SpO
1. Press the SETUP softkey.
2. Press the NEXT softkey three times.
Parameter Range
-
0
100%
-
0
250 bpm
-
0
255
varies from 0 to 100%,
2
value of 94.
2
N-600x Operator's Manual