Electrical

178
Accuracy Tolerance
1
Adult specifications are shown for
-
sensors with the N
600x. Neonate specifications are shown for
-
MAX
N sensors with the N
type. Refer to the Sensor Accuracy Grid shipped with the monitor.
The Sensor Accuracy Grid is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
2
Specification applies to N-600x monitor performance.
Reading accuracy in the presence of low perfusion (detected IR pulse
modulation amplitude 0.03% - 1.5%) was validated using signals
supplied by a patient simulator. SpO2 and pulse rate values were varied
across the monitoring range over a range of weak signal conditions and
compared to the known true saturation and pulse rate of the input
signals.
Display Update Interval
2 seconds
Instrument
Power
Rated at 100 to 120 volts AC (nominal 120 VAC) or
Requirements
220 to 240 volts AC (nominal 230 VAC), 20 volt/
amps to be compliant with IEC 60601
10.2.2.
Fuses
Quantity 2, 0.5 A, 250 volts, slow
IEC (5 x 20 mm).