Electrical; Environmental Conditions - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

Specifications

Electrical

Environmental
Conditions
76
Power Requirements
Fuses
The battery provides at least 2 hours of battery life when new and fully charged
with no alarms, no serial data, no analog output, no nurse call output, with
backlight on while using a pulse simulator set for 224 bpm, high light and low
modulation.
Type
Voltage
Recharge
Shelf Life
Complies With
Wavelength and
The wavelength range of the light emitted is within the range
Power
of 500 nm to 1,000 nm with the energy not exceeding 15
mW.
Temperature
Altitude/Barometric Pressure
Relative Humidity
Instrument
rated at 108 to 132 volts AC (nominal 120 VAC) or
200 to 240 volts AC (nominal 230 VAC), 20 volt/
amps to be compliant with IEC 60601-1 sub-clause
10.2.2
qty 2, 0.5 A, 250 volts, slow-blow, IEC (5 x 20 mm)
Battery
Lead acid
6 Volts DC
14 hours with N-595 turned off
18 hours with N-595 operating
2 months, new fully charged
battery
After 2 months storage the
N-595 will run for 50% of stated
battery life
91/157/EEC
Sensors
Operation
5 ºC to 40 ºC (41 ºF to 104 ºF)
-390 m to 3,658 m
(-1,280 ft. to 12,000 ft.)
70 kPa to 106 kPa
(31.3 in. Hg to 20.6 in. Hg)
15% to 95% non-condensing to be
compliant with IEC 60601-1,
sub-clause 44.5