Electrical

N-560 Operator's Manual
Power Requirements
100 to 240 volts AC, 25 to 39 volt/amps to be
compliant with IEC 60601
Fuses
qty 2, 2 A, 250 volts, slow
(5 x 20 mm)
The battery provides at least eight hours of battery life when new and fully
charged with no alarms, no serial data, while using a pulse simulator set for
60 bpm, high light and low modulation.
Type
Nickel metal hydride
Voltage
9.6 volts DC, 3.8 AH
Recharge
6 hours
Shelf Life
3 months, new, fully charged battery
After 3 months storage the N-560 will run for 50% of
stated battery life
Complies With
91/157/EEC
Electrical/Optical
Nellcor Pulse oximetry sensors contain light emitting
Specifications
diodes (LEDs) that emit red (~660 nm) and infrared
(~900 nm) light, with a total optical output power of less
than 15 mW. This information of sensor wavelength
range can be especially useful to clinicians, for example,
those performing photodynamic therapy.
Instrument
-
1 sub
-
blow, IEC
Battery
Sensors
Specifications
-
clause 10.2.2
99