Solving Problems and Updating Your System
Troubleshooting Resource Problems and COM Errors
Troubleshooting Resource Problems and COM
Resource Problems
CAUTION
It is possible to exceed Windows 95/98 allocated resource levels by opening
too many windows while running your Agilent E74xx system. This will cause
erratic behavior of the application and require the operator to reboot the
operating system in order to recover. This will also corrupt an open
Agilent E74xx database, with potential loss of data. Windows 95/98 will not
display any warning or feedback when this occurs.
The number of allowable open windows is dependent on the types you
choose, as different windows require varying amounts of Windows 95/98
resources. The maximum number of windows is approximately 12, although
up to 14 can be opened in certain configurations.
To provide the maximum amount of available Windows 95/98 resources to
the measurement application, it is strongly recommended that you close all
other applications before running the Agilent E74xx software. This includes
all applications found in the Startup folder. It is also recommended that you
run the Windows 95/98 resource meter in a minimized configuration in order
to monitor available remaining resources.
To start the Windows 95/98 resource meter:
1. Choose Start > Programs > Accessories > System Tools > Resource
Meter. The Resource Meter dialog box appears.
2. Click
While this application utilizes system resources, it provides a way of
determining when the system is running low on available resources. Opening
Agilent Wireless Solutions
Errors
.
OK
83