MULTIMOBIL 2.5
Maintenance Instructions
Print No.: SPR8-X01.831.01.02.02
Replaces: SPR8-X01.831.01.01.02
SP
© Siemens AG 2003
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 09.05

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MULTIMOBIL 2.5

  • Page 1 MULTIMOBIL 2.5 Maintenance Instructions © Siemens AG 2003 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility...
  • Page 2 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. Siemens Ltd. Med India Version 6.0 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 3: Table Of Contents

    Removing the X-ray source assembly..............8-9 8.11 .................... 8-10 EPLACING THE SINGLE TANK 8.12 .................... 8-10 EPLACING THE COLLIMATOR 8.13 ..................8-11 EPLACING THE OLLIMATOR LAMP Siemens Ltd. Med India Version 6.0 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 4: Maintenance Instructions

    Replacing damaged or missing screws Before working on the open MULTIMOBIL 2.5 switch off the unit with the power switch on the control panel and unplug the power cable. Before inserting or removing PC boards, observe the ESD regulations.
  • Page 5: Product Disposal

    Legal regulations may contain special prescriptions concerning the disposal of this product. In order to avoid environmental damage and/or personal injury, please inform SIEMENS Customer Service if you want to put the unit out of operation and dispose it. 8.4.1 Radioprotection Material Lead in X-ray tube assembly and in Collimator 1.65 kg.
  • Page 6: Safety Checks

    Check function of locking lever and cap screw on the X-ray tube. Check the legibility of the labels according to those in section Location of Labels. If any label requires to be replaced, please notify Siemens customer service. Performance Check (Every Six months) Check Counterbalancing of the support arm at each position.
  • Page 7: Aligning The Light Field To The Radiation Field

    Repeat the check and, if necessary, adjust the collimator again until the deviation between the light field and the radiation field is within the tolerance (≤ 2 cm). Siemens Ltd. Med India Version 6.0 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 8: Checking And Readjusting The Counterweight

    If it moves UP, loosen the M10 nut. Check the balancing and then tighten the locknut. Siemens Ltd. Med India Version 6.0 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 9: Removing The Top Panel , Front Cover And X-Ray Source Assembly

    8.10 Removing the top panel , front cover and X-ray source assembly 8.10.1 Removing the top panel Turn off the Multimobil 2.5,disconnect the mains chord and wind the cord around the cable holder. Lower the X-ray source assembly to transport position. Lock the Multimobil 2.5 by applying the brake.
  • Page 10: Removing The

    Position the front cover on a soft bedding. 8.10.3 Removing the X-ray source assembly. Turn off the Multimobil 2.5,disconnect the mains chord and wind the cord around the cable holder. Lower the X-ray source assembly to transport position. Lock the movement of the Multimobil 2.5 by applying the foot brake.
  • Page 11: Replacing The Single Tank

    Proceed as described in the following Loosen the allied connector on the collimator cable and disconnect it. Remove the collimator after loosening the four screws . Siemens Ltd. Med India Version 6.0 8-10 Copyright © SIEMENS LTD. All rights reserved. For internal use only...
  • Page 12: Replacing The Collimator Lamp

    Do not touch the glass envelope with your CAUTION bare fingers. Screw down the bracket and the front panel.Check the function of the lamp. Siemens Ltd. Med India Version 6.0 8-11 Copyright © SIEMENS LTD. All rights reserved. For internal use only...