Installing the ON/OFF assembly
Wiring of terminals D50.X2, D50.X11 and D50.X4
WARNING
•
UPS power input: Disconnect the power cable with the installed power coupler from the
old ON/OFF assembly and use it for wiring the UPS according to the table.
•
UPS power output: Use the power cable with installed power plug included in the
supplied cable set for wiring the UPS according to the table.
•
Connect the power cables of the components installed in the monitor trolley according to
the table:
Component
Cable
blue
Cable
brown
Cable
yellow/green
(ground wire)
Wiring of terminal X3
•
Connect cables X3.1 to X3.6 (monitor trolley cable) according to the following table:
Color of monitor trolley cable green
Terminal D50
Installing the ON/OFF assembly
Insert the ON/OFF assembly into the monitor trolley and fasten it with the 4 Allen screws.
In doing so, ensure that there is no mechanical strain on the cables and that they are not
pinched.
Siemens AG
Medical Solutions
Electrical safety!
Non-compliance can lead to severe injuries and even death as
well as to material damage.
Observe the safety information in chapter 1 of these instructions.
Do not yet connect the UPS input and output power cables to
the UPS!
Terminals on the ON/OFF assembly D50
Live
Reference
(left)
(right)
monitor
monitor
X11.2
X2.1
X11.5
X2.2
PE
PE
(ground wire
(ground wire
terminal)
terminal)
X3.1
n.a.
SP00-000.841.01
Rev. 01
PC
UPS
(image
input
processor)
X11.3
X4.2
X11.6
X4.3
PE
PE
(ground wire
(ground wire
terminal)
terminal)
red
yellow
black
X3.2
X3.3
X3.4
Page 5 of 10
08.05
CS PS 24
3 - 5
UPS
Options
output
(e.g. int.
printer)
X11.1
X2.1
X11.4
X2.2
PE
PE
(ground wire
(ground wire
terminal)
terminal)
violet
blue
X3.5
X3.6
ARCADIS
System Manual
3
3