Test 3: Detector Uniformity

Objective

To measure the uniformity of the detector response over its entire surface

Required Equipment

a)
Collimator mountable PMMA phantom
b)
Compression plate 24 x 30 cm

Procedure

1.
Create a new patient record in the local database.
Fill in: Last Name:
2.
Select procedure QC-raw.
3.
Mount PMMA phantom close to collimator.
4.
Mount compression plate and perform compression.
5.
Select exposure technique AEC
28 kV
target/filter combination Mo/Mo
grid in
auto decompression off
OPCOMP off
AEC segmentation off
low dose mode off
Change all projection views to L-CC
6.
Perform exposure.
MAMMOMAT Inspiration
SPB7-330.640.01.01.02
Test_Three
First Name: Detector Uniformity
Patient ID:
day+time when the test is performed
(example: 200707071030)
DOB:
05 05 1955
Gender:
Other
Test 3: Detector Uniformity
2
23