MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
Print No. MR-04010G.623.22.02.02
Positioning the female patient
4.1.3
1
Position the female patient in the prone position with the breasts
in the recesses.
2
When positioning the patient with the arms above the head:
Position a suitable pad between the hands to avoid current loops.
3
Use suitable positioning aids.
Performing the measurement
4.1.4
1
Align the laser-light localizer with the center marking of the coil.
2
Start the measurement.
4.2
Quality measurement
The quality measurement of the 4 Ch BI Breast is performed in the
head first and the feet first orientations.
Use the following accessories for the quality measurement:
◾
MAGNETOM Avanto: 2000 ml plastic bottle
◾
MAGNETOM Verio: 1900 ml plastic bottle + 3 cm spacer cushion
✓
All RF coils have been removed from the patient table.
✓
MAGNETOM Verio only: The appropriate spacer cushion is
positioned on the patient table.
✓
4 Ch BI Breast is positioned on the frame and connected.
Breast imaging
4
27