Required Tools, Measurement And Auxiliary Devices - Siemens MAMMOMAT 1000 Nova Opdima Maintenance Instruction

Hide thumbs Also See for MAMMOMAT 1000 Nova Opdima:
Table of Contents

Advertisement

6
1.3

Required tools, measurement and auxiliary devices

NOTE
Protective ground wire tester (44 15 899 RV090)
Stereo calibration phantom (included in the Opdima® delivery) (part no. 64 30 701)
Acrylic plastic (PMMA = Poly-methyl-met-acrylate) plates: 3 plates of AEC calibration
Plexiglas 150 mm x 150 mm x 19 mm (part no. 65 61 232) and one plate of calibration
Plexiglas, 150 mm x 150 mm x 9.7 mm (part mo. 65 61 224)
Grid object table
Mammography bar pattern (5-20 lp/mm), for instance Nuclear Associates Resolution
Test Pattern (07-555), http://www.nucl.com
Cleaning agents* specified in the Instructions for Use MAMMOMAT 1000/3000 Nova,
chapter "Cleaning, disinfection"
Standard service tools*
MAMMOMAT Nova - Opdima SPB7-230.101.03.05.02
All listed tools, measurement and auxiliary devices, with the ex-
ception of those marked with "*", are specified in the STC (Service
Tools Catalog).
02.06
General information
Page 6 of 18
CS PS SP
Siemens AG
Medical Solutions