Refilling oil-tank
N
Follow this procedure to refill the oil-tank:
N
Unplug the device and wait for one minute.
N
The oil-tank is located in the external control unit of the device below at the left side (At device up to the
construction year 2002 and without external control unit the oil-tank is located at the rear position of the
machine close to the drive engine.).
N
Open the lid of the tank and refill the oil-tank with approximately 0.5 litre (at devices up to construction year
2002 and without external control unit approximately 1 to 1.5 litre) with the provided original special oil, to
receive at h/p/cosmos.
N
Close the lid of the oil-tank.
N
Plug in the power plug of the device and switch device on.
N
Check control lamp at the external control unit. If it is not indicating anymore, the oil-tank has been correctly
refilled.
N
If necessary quit oil-message with „OPTION 01".
Special oil for sliding plate with automatic lubrication/0.25l-bottle [cos14007]
Special oil for sliding plate with automatic lubrication/0.50l-bottle [cos13438]
Set back of oil message MCU 4 without UserTerminal
[7.D5]
An external UserTerminal or the PC-software h/p/cosmos para control
"Devices with UserTerminal".
Set back of oil message MCU with UserTerminal
[7.D6]
Start: Running belt is not moving. One of the mode LED ~ of operation is flashing: (manual, profile, cardio, test)
Step
Activity
[7.D5]
[01]
Select of the mode for "user
options" (OPxx)
[02]
Confirm option no. 1
[03]
Delete message
[04]
Quit option mode
© 2008 h/p/cosmos sports & medical gmbh
Keys
a d b
all 3 keys
simultaneously for
at least 3 seconds
c
c
b
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
maintenance & safety inspections
®
is required for this. Further instructions see
Response/Display
=
indicates: OP01 flashing (for option no. 01)
!
"
and
indicate: E.rE SEt (for error reset)
§
indicates: donE to inform you, that you have to
delete the oil message
=
indicates: OP01 flashing (for option no. 01)
!
"
and
indicate: E.rE SEt (or error reset)
Stand-by mode
~
manual, profil, cardio or test is flashing
created 01.09.2008
printed 08.10.2008