[05]
[06]
[07]
[08]
[09]
© 2008 h/p/cosmos sports & medical gmbh
Immediately at first installation an electric safety test / measurment
has to be performed for „Protective Earth Resistance", „Isolation
Resistance" and „Leakage Current" and the values have to be
recorded on a special protocol [cos11690xx] and marked with „first
measured values". For details please refer to chapter
„maintenance / inspections". One copy of this protocol
[cos11690xx] remains at the owner's manual and the original with
the „first measured values" shall be sent to h/p/cosmos. The
potential equalization cable has to be
connected AFTER the measurements.
Lubricate the running machine before introduction according to
instruction with 30 ml special silicon oil which is part of the "service
box".
Check the tension of the running belt (no slip).
Adjust the running belt properly before instruction with the
customer.
Test the function of the heart rate transmission with the help of a
simulator or a POLAR sender. Search for interferences and
possible reasons.
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
installation & commissioning
The allen-key must be pulled out of the screw
immediately after usage due to risk of injury!
created 01.09.2008
printed 08.10.2008
accomplished
accomplished
accomplished
accomplished
accomplished
page: 37 of 176