h/p/cosmos 150/50 locomotion de med Operating Instructions And Service Manual page 58

Hide thumbs Also See for 150/50 locomotion de med:
Table of Contents

Advertisement

The „Gardner Testprotocol"
11
[5.B13]
for application Angiology.
The Gardner Testprotocol serves for evaluation of
the maximum walking distance of Peripheral
Arterial
Claudication.
The test has to be performed under permanent
supervision of a doctor and a fall prevention
system (e.g. safety arch with chest belt) has to be
used.
The patient first stands on the side-footboards of
the running machine and not on the belt. Start test
profile 11 and the belt speeds up to 3.2 km/h. As
the patient steps on the running belt the doctor
press the START key again and by pressing the
START key the second time the displays will be
reset to zero.
After completing the test the results can be printed
on a host printer, if connected to the interface of
the running machine directly.
12 - 20
not engaged / reserved
21 - 28
free definable user profiles
Additional profiles in the test mode from firmware 3.02.3 – only optionally selectable, locked by default
70
Training
[5.B13]
Interval 600 m / 14.5 ... 17.5 km/h
© 2008 h/p/cosmos sports & medical gmbh
Disease
Patients
with
Intermittent
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
Step
Duration
Speed
min:sec
Pre-Testphase. Patient standing on the footboards, not on the
belt
0
until
3.2 km/h
START
is pressed
Testphase. Patient steps onto the running belt.
1
02:00
3.2 km/h
2
02:00
3.2 km/h
3
02:00
3.2 km/h
4
02:00
3.2 km/h
5
02:00
3.2 km/h
6
02:00
3.2 km/h
7
02:00
3.2 km/h
8
02:00
3.2 km/h
9
02:00
3.2 km/h
10
02:00
3.2 km/h
11
30:00
3.2 km/h
for following extensions and up-dates
max. 40 Program steps / not scaleable
Step
Distance/time
Speed
1
600 m
14.5 km/h
2
1:30 min
0.0 km/h
3
600 m
15.5 km/h
4
1:30 min
0.0 km/h
5
600 m
16.5 km/h
6
1:30 min
0.0 km/h
7
600 m
17.5 km/h
8
1:30 min
0.0 km/h
created 01.09.2008
printed 08.10.2008
operation
Elevation
Total Time
min:sec
0
until START
is pressed
0
2:00
2
4:00
4
6:00
6
8:00
8
10:00
10
12:00
12
14:00
14
16:00
16
18:00
18
20:00
18
50:00
Elevation
Acceleration
0 %