Oiling amount
3 syringes each filled with 10 ml silicone oil. (special devices might have different oiling amounts - see delivery
papers) and 1 syringe filled with air (to empty the oil line in the running machine).
Procedure
N
Start the running machine at a max. speed of 5 km/h.
N
Fill the syringe with 10ml of the lubricant.
N
Launch the syringe into the oil-filling hole and press the
contents slowly in.
N
Fill the syringe again with the lubricant and slowly press
the contents into the oil-filling hole.
N
Fill the syringe again with the lubricant and slowly press
the contents into the oil-filling hole.
N
Fill the syringe again with air and slowly press the air
into the oil-filling hole. This is to empty the small oil-
tube of the running machine.
N
Clean the area around the oil-filling-hole in order to make it free of overrunned oil.
N
Let the running machine operate for a few minutes at the speed of 5 km/h, so that the oil can spread itself on
the inside of the running belt. Walk for 2 minutes at very low speed (ca. 2 km/h) and vary your positions on the
running belt, so that the oil under the belt is spread evenly.
N
Check the position of the running belt after the lubrication. The belt should be in a centre position on the rear
roller (tail shaft). In case the belt needs to be adjusted please follow the separate instruction.
N
After lubrication the oil message has to be terminated with "OPTION 01" (see optional functions). The device
has no sensor and the oil message will not terminate automatically after lubrication. Refer to the procedure
described at „Set back of oil-message" in this chapter.
This oil type must not be used for running-decks for cycle-
and wheelchair - application (e.g. saturn 300/100r), and
not for oil-tanks with automatic oil-pump.
Special oil 0.25 l - bottle
for sliding plate with manual lubrication
[cos000200008001]
Syringe 10 ml for lubrication of sliding plate
[cos12181]
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
© 2008 h/p/cosmos sports & medical gmbh
email@h-p-cosmos.com
created 01.09.2008
maintenance & safety inspections
printed 08.10.2008
page: 105 of 176