Standard settings for the user options
[6.A3]
Option
Character / Function of option
[6.A3]
OP 01
Set back of error messages
OP 02
Total distance (km)
OP 03
Total hours of operation stand-by incl. runtime (h)
OP 04
Total hours of operation runtime only (h)
OP 05
Firmware Version
OP 06
Adjustment of the real time clock
OP 07
Acoustic Heart-Rate-signal
OP 08
Stopping / Deceleration Time
OP 09
Starting Speed (Modus Manual and Cardio)
OP 11
Scaling of Profile mode
OP 12
Unit for display of speed – one decimal place rsp. none
for 3 = m/min
Unit for display of speed – two decimal places rsp.
none for 3 = m/min
OP 13
Unit for display of distance
OP 14
Unit for angle of elevation
OP 15
Subject's Body Weight
OP 16
Request for body weight before manual or automatic
start
OP 17
Unit for energy consumption
OP 18
Maximum speed (default) in Cardio mode
OP19
Setting of Polar W:I:N:D System
OP 20
RS 232 interface protocol: COM 1
OP 21
RS 232 interface protocol: COM 2
OP 23
RS 232 interface protocol: COM 4
OP 27
Minimum acceleration and deceleration level
OP 28
Maximum acceleration and deceleration level
OP 29
Acceleration and deceleration level for remote control
via RS232 interface
OP 40
Locking and unlocking the treadmill
OP 41
Locking and unlocking the Manual mode
OP 42
Locking and unlocking the Profile mode
OP 43
Locking and unlocking the Cardio mode
OP 44
Locking and unlocking the Test mode
OP 45
Report mode Display „index"
OP 46
Report mode Display „elevation"
OP47
Sustain values in display resp. automatic „Reset"
OP48
Count-Down of program-step
OP 52
Output interval for printer protocol
OP 53
Language settings for printer protocol
© 2008 h/p/cosmos sports & medical gmbh
.
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
Default setting
Report only
Report only
Report only
Report only
current date/time
OFF
5 seconds
0.5 km/h
0
0 = km/h
2: m
0: % (Prozent)
65 kg
OFF (no request)
JOUL = kJoule
6.0 km/h
0000 0000
1 = h/p/cosmos coscom
1 = h/p/cosmos coscom
20 = h/p/cosmos coscom v3
1
4
1
ON (unlocked)
ON (unlocked)
6 (unlocked up to profile 6)
ON (unlocked)
28 (unlocked up to test 28)
0
0
OFF
OFF
60 (seconds)
English
created 01.09.2008
printed 08.10.2008
optional settings
Adjustable range
0 ... 4.294.976 km
0 ... 1.193.046 h
0 ... 1.193.046 h
x.xx.x
currently ... 31.12.2092
OFF or ON
2 ... 30 seconds
0.0 km/h ... 5.0 km/h
0. 1 or 2
0 = km/h 1 = m/s
2 = mph 3 = m/min
20 = km/h 21 = m/s
22 = mph 23 = m/min
0: km;1: miles; 2: m
0 = % / 1 = ° (Grad)
10 ... 250 kg
OFF / ON
JOUL = kJoule
CALO = kcal
0.0 ... max. speed
xxxx xxxx
1 ... 20
1 ... 18
OFF,. 18,. 20, 22
1 ... 5
1 ... 7
1 ... 5
OFF = locked
ON = unlocked
OFF = locked
ON = unlocked
0 ... 6
OFF = locked
ON = unlocked
0 ... 94
0 ... 3
0 ... 2
OFF = RESET automat.
ON= RESET 2xSTOP
OFF = count-up
ON = count-down
0 = no single values,
1 ... 100
English, German,
French, Spanish
Portuguese, Hungarian
page: 82 of 176