Activating and de-activating motor break The motor break prevents the running belt from manual moving (for
[6.B1]
OP14
Maximum acceleration time for RS232
(option only available for treadmills, not
for ladder ergometers)
OP16
Activating and de-activating heart rate
measurement
OP33
Distance between rungs of the ladder
ergometer
(option
ergometers, not for treadmills)
© 2008 h/p/cosmos sports & medical gmbh
only
available
for
ladder
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
example because of the subjects bodyweight) when the speed value is
set to "0". If the treadmill recognizes a running belt movement it
activates the motor break for 10 seconds.
.
Remark:
The motor break is only working at running machines with integrated
speed measurement. The speed sensor must be switched on
(OP11=1). If the running machine is not equipped with speed
measuring system this could be retrofitted easily.
To activate the motor break it is necessary to additionally change
some parameters at the frequency inverter. Ask the service
department of h/p/cosmos sports & medical gmbh for detailed
instruction.
Attention!
If the device is switched off or disconnected from
the power supply the motor break is not working!
After 10 seconds of operation of the motor break,
the break is released automatically and is only
activated again after a belt movement. A short jerk
may occure during reactivation.
§
shows OFF = Motor break not active
§
shows ON = Motor break active
§
shows: e.g. 5 (max. acceleration-increment)
$
shows: e.g. 131.1 (max. acceleration-time in sec.)
%
SEC
shows:
.
=
reports ...
0 = heart rate measurement is de-activated / OFF
1 = heart rate measurement is activated, circuit board HFU1 or
RE06REC (default)
2 = heart rate measurement is activated, circuit board HFUi
auto = automatic detection and activation of circuit board
4 = Polar W:I:N:D (further settings in user OP19 and OP23 necessary)
For correct calculation of the number of rungs / steps, the distance
between two rungs can be changed and confirmed in this option.
Standard: 254 for 25,4 cm
created 01.09.2008
printed 08.10.2008
optional settings