N
After having switched on a 3 phase machine, an elevation of 0 % has to be adjusted automatically. If this is not the
case, the unit is to be switched off immediately and the two phases of the wall socket have to be changed so that
the device is supplied with right-handed polarized power. Otherwise the elevation cannot be set correctly and the
power supply of the motor for controlling the elevation is switched off via the limit stop switch at the ground frame.
N
Use a usual 16 Ampere expulsion fuse with C-tripping-circuit for your house distribution. For running machine-
models with a construction size of the running surface of 200/75cm up to 300/125cm you will need a 32 Ampere
expulsion fuse with C-tripping-circuit. If however, the expulsion fuses switch off when starting the device, the circuit
has to be secured with a blowout fuse or an expulsion fuse with another tripping-circuit (e.g.: K–fuse). Read name
plate and technical data of the machine just to make sure.
N
For further questions please ask your electrical engineer or h/p/cosmos.
N
Before installing the running machine please compare the specifications on the nameplate concerning the mains
voltage and the mains frequency with your local characteristics. Connection only if identical.
N
Check the main lead, the voltage power supply outlet and ground wire protection-contacts before plugging it in.
Damaged leads and couplers and defective or dirty contacts have to be exchanged immediately. Rubber-leads can
get porous and friable after some years.
N
Plug the running machines directly into the wall socket. Each running machine has to be
connected to a separate circuit. The socket has to be marked with name and serial number of
the running machine. The use of an extension cables or a multiple plug sockets is not
allowed.
N
Electrical devices with mains connections must neither be used in wet and humid areas (e.g.
swimming pool, sauna, etc.) nor in environmental chambers.
Name plate
[4.C1]
Name plate examples sports device model h/p/cosmos quasar / medical device h/p/cosmos saturn 300/125r
For full details please refer to the chapter „technical data" in this manual.
© 2008 h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
installation & commissioning
created 01.09.2008
printed 08.10.2008
page: 27 of 176