h/p/cosmos 150/50 locomotion de med Operating Instructions And Service Manual page 87

Hide thumbs Also See for 150/50 locomotion de med:
Table of Contents

Advertisement

OP34
Manual speed calibration
[6.B1]
(counted increments from the speed
sensor for 10 m belt movement)
© 2008 h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
Normally the speed calibration values are set by selecting the correct
device type (administrator option number 3). Only for special speed
and special models the speed calibration values have to be changed. If
you change the speed calibration value, please also pay attention to
the correct settings in option number 5 (firmware speed limit), number
11 (speed measurement; must be switched on) and option number 48
(Frequency inverter speed limit) of the administrator mode. To change
the speed calibration value please use the + and – buttons. For quick
movements between standard values, please use the UP- and DOWN-
buttons.
If the speed measurement is switched off in OP11, then OP34 will only
report internal program absolutes. In the other case following values
are stored:
N
Running machine series h/p/cosmos mercury with
special speed 10 km/h:
Speed Calibration Value: 8.490
Inverter Frequency:
N
Running machine series h/p/cosmos mercury with
speed 22 km/h:
Speed Calibration Value: 5.988
Inverter Frequency:
N
Running machine series h/p/cosmos mercury with
special speed 30 km/h:
Speed Calibration Value: 4.165
Inverter Frequency:
N
Running machine series h/p/cosmos quasar with
special speed 10 km/h:
Speed Calibration Value: 7.800
Inverter Frequency:
N
Running machine series h/p/cosmos quasar with
special speed 22 km/h:
Speed Calibration Value: 4.495
Inverter Frequency:
N
Running machine series h/p/cosmos quasar with
speed 25 km/h:
Speed Calibration Value: 4.495
Inverter Frequency:
N
Running machine series h/p/cosmos quasar with
special speed 30 km/h:
Speed Calibration Value: 4.010
Inverter Frequency:
N
Running machine series h/p/cosmos quasar with
special speed 40 km/h:
Speed Calibration Value: 2.874
Inverter Frequency:
Continuation next page...
created 01.09.2008
printed 08.10.2008
optional settings
80.2 Hz
123.3 Hz
119.0 Hz
80.0 Hz
94.5 Hz
110.0 Hz
115.0 Hz
113 Hz
page: 87 of 176