OP39
Elevation: display as set point or true
[6.B1]
value
(This option is only available for
treadmills, not for a ladder-ergometer)
OP40
Speed: display as set point or true value
OP41
Elevation control after STOP
(This option is only available for
treadmills, not for a ladder-ergometer).
OP46
Display and setting of elevation for
ladder ergometer
(only available for ladder ergometers,
not for treadmills)
OP47
Set back of oil- and service interval
© 2008 h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
0 = Display of elevation as set point
1 = Display of elevation as true value rsp. flashing set point report, if
the set point was changed with the keys and has not yet been
reached. (Report: current elevation angle; recommended default-
setting)
0 = Display of speed as set point (Report: target speed)
1 = Display of speed as true value rsp. flashing set point report, if the
set point was changed with the keys and has not yet been reached.
(Report: current speed; recommended default-setting)
0 = Elevation maintains at the present level after pressing the STOP
button (recommended; default-setting)
1 = Elevation gets down to 0 % after pressing the STOP button.
The setting of option 41 to value 1 effects the
safety of the subject and of third parties. A down-
movement of the elevation after pressing STOP
can result in a jumpy reaction and in dangerous
injuries and contusions.
h/p/cosmos strongly recommends to keep the
standard setting of this option. Stop means Stop
for all movements and must not result in an active
action of the running machine. It is only allowed
to change option 41 after a written documented
instruction about the danger warnings and after a
written confirmation of the customer/user.
$
indicates: Elevation of ladder ergometer, standard: 75 degree,
adjustable from 0 to 180 degree
After a service with lubricating the treadmill and performing all
maintenance work the oil and service interval can be set back with
option number OP47.
created 01.09.2008
printed 08.10.2008
optional settings
page: 90 of 176