OP49
Type of frequency inverter (FI) control
[6.B1]
OP51
Error statistics of the digital FI control
© 2008 h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1.04hpc-en\20080901_cos14310m5-v1.04hpc-en_manual_h-p-cosmos_running_machine.doc
email@h-p-cosmos.com
Setting of FI control. Two types are available for selection:
§
Analogue 0 to 10 V
Digital via RS-485
The lower display row reports the occurred errors of three different
error types.
§
Number of NAK errors: the FI has sent an NAK (not
acknowledged), i.e. the FI has not accepted the last command
(transmission error, ...).
$
Number of proof total errors: the proof total of the received FI
messages is wrong (transmission error, ...).
%
Number of Timeout errors: the FI has not answered within the
Timeout period (60 ms).
Note: At each emergency-stop the number of timeout errors is
increased compulsorily by at least one numerator, because the
transmission line to the FI is being interrupted by hardware.
created 01.09.2008
printed 08.10.2008
optional settings
$
%
AnA
LOG
rS-
485
page: 92 of 176