Microscan QX Hawk User Manual page 288

Industrial imager
Hide thumbs Also See for QX Hawk:
Table of Contents

Advertisement

Default/Reset/Save
Customer Default Parameters
Customer default parameters (saved by <Zc>) are the same set of parameters as power-on
parameters but are saved in a different, isolated section of NOVRAM. This allows a user
essentially to create a back-up set of parameters that can be recalled in the event that the
current parameters or power-on parameters have been accidentally changed or are no
longer desired.
It is important to note that a hardware default does not affect customer default parameters.
For example, a user that has inadvertently changed communication settings and saved
them with a
<Z>
command, may not know the correct settings or have the capability to
communicate within those settings. By first doing an EZ Button default to restore the
known Microscan defaults, the user can then recall the previous customer-saved settings
with an
<Arc>
or
<Zrc>
Microscan Default Parameters
Microscan default parameters are contained in the firmware and cannot be changed.
Software Defaults
Microscan default parameters can be recalled (loaded into current settings) with the
<Ard>
command or recalled and saved for power-on with the
Hardware Default
If a software default and reset is not possible, it may be necessary to reset the imager by
shorting (connecting) specific pins. This procedure has the same effect as the
software command.
Important: For this reset to occur, this command must be executed within 60 seconds of a
power-on or a reset.
1. Apply power to the imager.
2. If using an IB-131, locate pins 7 and 11 on the host connector.
Caution: Be certain that the correct pins are located. Connecting the wrong pins could
cause serious damage to the unit.
3. Momentarily connect these wires (or pins) and listen for a series of short beeps.
4. Within 3 seconds, connect them again. A longer beep should be heard. If not, repeat
the process.
Default on Power-On
Use the EZ Button to default the imager by holding it down while applying power to the
imager (if this feature is enabled).