Microscan QX Hawk User Manual page 142

Industrial imager
Hide thumbs Also See for QX Hawk:
Table of Contents

Advertisement

Read Cycle
Time Between Captures 7-8
This value specifies the delay to wait before starting a frame capture for the next capture
in the rapid capture sequence.
Notes for Rapid Capture, Timed Capture
If rapid capture is configured for timed captures this delay is measured from the point at
which the previous delay expired. If the delay expires during a previous capture the start of
the capture will occur as soon as the previous capture or captures complete. For example,
if rapid capture is configured for 8 captures and all delays are configured for zero all 8 captures
will occur back to back as fast as the imager is able.
Notes for Rapid Capture, Triggered Capture
If rapid capture is configured for triggered captures mode the delay taken is measured
from the point at which the trigger for that capture is received. Also if an incoming trigger
occurs before a previous delay has expired, the previous capture is initiated and the delay
for the incoming trigger is started.
Important: If the unit is triggered to capture more than 8 images the unit will repeat the
last delay (8th) for the remaining frames.
5-58
QX Hawk Industrial Imager User's Manual