Pressing the
Pressing the
mode without printing the report.
If during the control course the number of negative T errors detected by the
software requires controlling another sample from the batch (as specified by the
regulation on CPG), then on completing the measuring process of sample 1 the
software displays a command to control another sample from the batch:
Accept the message by pressing
quantities visible in the control window change. On completing control of the
second sample the software generates a summary (report) from the process and
enables printing a report on a connected printer.
Caution:
A template and an example of a report from product control is provided in point
26.7 of this user manual.
A template and an example of a report from determining average tare of a
packaging is described in point 26.6 of this user manual.

26.4. Non-destructive empty-full control mode

In the "Non-destructive empty-full" control in the product settings the user
sets a measuring "lot" (manufacturing unit). The software, according to the set
"lot" displays a command ordering first weighing on an empty packages, and
then to weigh the same packages after filling with maintaining their weighing
sequence.
Caution:
Editing products in the database of packaging in described in point 27.2 of this
user manual.
causes printing a report on a printer connected to the balance.
key causes returning to the main window of the <
key. The texts and the permissible error