Additional Settings Of Density Mode - RADWAG UYA 3Y User Manual

Balances 3y series
Hide thumbs Also See for UYA 3Y:
Table of Contents

Advertisement

9. Load aluminum mass standard on balance's weighing pan and on
stabilization of measurement result accept it by pressing <
10. The balance automatically calculates the density of air, and indicates its
value on the display,
11. Press <
The value of determined air density is automatically assigned to field <Air
density> in menu <Settings/ Air buoyancy compensation> in working mode:
weighing.

20.4. Additional settings of Density mode

The additional settings enable adjusting the working mode to user's needs and
requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout
and Profile,
3. Press <Settings> key,
4. The display indicates functions related to density determining
process, such as:
o Request sample number
Available settings:
NO – data on sample number is not required; applicable in case of
a single measurement
YES – before each measurement a window requesting inserting
sample number is automatically displayed; applicable in case of a
series of measurements
o Measuring unit: the balance enables choosing between the
following units: [g/cm
for all density determining options and also in final printout.
o Assign density to a product: YES/NO – if option <YES> is
selected, the software automatically assigns the value of
determined density of solid or liquid to field <Density> in a
> key to finish the procedure.
3
3
], [kg/m
], [g/l]. Selected unit is activated
- 76 -
> key,

Advertisement

Table of Contents
loading

Table of Contents