Hide thumbs Also See for UYA 3Y:
Table of Contents

Advertisement

3. Press <Settings> key,
4. The display indicates functions related to calibration of pipettes.
Pipettes calibration is related to the following functions:
− Number of measurements; this option enables determining number of
measurements for tested volume of a pipette (valid for each tested
volume in case of a pipette with adjustable volume),
− Request serial number: YES/NO. For option <YES> before starting
calibration procedure, the software displays a window and on-screen
keyboard for inserting serial number of a pipette,
− Comply with ISO 8655: YES/NO. For option <YES>, for tested
volumes, during pipettes calibration process the software automatically
accepts error values as specified in standard ISO 8655 (if other error
values have been accepted for a pipette, then on selecting option
<Comply with ISO 8655> the error values are neglected, and replaced by
those from ISO standard),
− Acquire data on ambient conditions from THB module: YES/NO. For
option <YES> the software automatically, and in specified moment,
acquires and saves ambient conditions' parameters from a connected THB
module. If option <NO> is selected, before starting and after completing
the pipettes calibration process operator should manually insert ambient
conditions parameters (i.e. temperature, humidity and pressure in the
weighing room) from external ambient conditions sensors.
− Result control, Tare mode, Automatic footer printout, Printout
mode, Printout, means of using these functions is given in point 10.8.
"Additional parameters on weighing process".
24.2. Pipettes calibration – quick access keys
Each of working modes features a set of default quick access keys which are
automatically displayed on mode activation. The set of keys can be modified
by selecting other quick access keys to the bottom bar of the display. Such
process requires appropriate operator's access level.

24.3. Adding a pipette to the Database of Pipettes

Database of pipettes contains list of pipettes' names and other data including
tested volume and errors values for a specific volume. When adding a pipette
to the database, first specify its name, and then add other data on a pipette.
The software operates intuitively, and commands a user to follow displayed
information.
Adding a pipette to the Database can be carried out in menu of the database.
- 99 -