10.6. Weighing on dual range balances (ref. balance PS 200/2000.3Y
series)
Transition from weighing with accuracy of the
with accuracy of the
exceeding Max
200g (without operator's interference). On entering the
1
accuracy of the II weighing range, the display indicates <
left side, the last digit is simultaneously blanked, and last but one digit is
differentiated.
From now on the balance continues weighing in the II weighing range.
In order to return to weighing with the accuracy of the
follow below procedure:
• take the load off balance's weighing pan;
• as the indication returns to zero and symbols: →0← and
press →0← key.
The balance returns to weighing with readability d
II weighing range <

10.7. Use of air buoyancy correction factor

The application enables correcting mass measurement errors during:
1. Weighing loads that density considerably differs from the density of
standard used to adjusting the balance. Regularly, the balance is adjusted
with a steel mass standard with density ~8.0g/cm
standard with density ~8.7g/cm
weighed, then ratio indicated on below chart should apply.
Below chart presents the value of mass correction in relation to the density
of weighed load, assuming that air density is constant and equals
3
1.2 kg/m
.
weighing range takes place automatically on
II
> is blanked.
II
3
. If loads made of other substances are
- 31 -
weighing range to weighing
I
> pictogram on its
II
I weighing range
II weighing range
weighing range
I
=0.001g and the marker of
1
3
or a brass mass
are lit up,