mass of an ingredient if it is within set percent tolerance of total
ingredient's mass (±%) - (data acquired from database of products).
THRESHOLDS: the software accepts mass of an ingredient as correct if
it is within set thresholds (data acquired from database of products).
Selected relation is in effect for all ingredients in the formula making
process.
− Multiplier editing, enables making multiple mixtures according to a
selected formulation in a single weighing process:
o YES, on selecting a formulation for processing, the software
requests for specifying multiplier's value used for multiplying mass
of each ingredient while weighing. Inserted value is in effect for all
ingredients in the formula making process.
o NO, inserting multiplier value is disabled, default value is set to
<1>
− Printout, means of using the function is given in point 10.8. "Additional
parameters on weighing process".
22.2. Formulation – quick access keys
Each of working modes features a set of default quick access keys which are
automatically displayed on mode activation. The set of keys can be modified
by selecting other quick access keys to the bottom bar of the display. Such
process requires appropriate operator's access level.

22.3. Adding formulas to the Database of Formulas

Database of formulas consists of formulas names, and ingredients with names
and mass forming a formulation. Each product used for formula making
process is saved in the database of products. When preparing a formulation,
the first step requires naming the formulation, and then adding ingredient's to
already named formulation.
The software operates intuitively, and commands a user to follow displayed
information.
As each of ingredients must have its name and mass specified, the operator
must know the exact composition of the total mixture. Adding a formulation to
the database of formulas is carried out from the database level.
Procedure:
Enter submenu <
Press <
Press:
– to add a new formulation to the database
– to abort the process
Databases>, press < Formulation> field
Add > soft key, should a formulation be added