Empty - Full Control - RADWAG Y/KTPS Series User Manual

Wly/ktps series; wpy/ktps series control of packaged goods network version
Table of Contents

Advertisement

[1] Product test
Qn
TARE
PUT 1 / 30
T
2T
Th. 1
av.
OK
Test in progress
F5-End
NET
• According to the command on the screen put on the weight pan
subsequent products and after stabilizing press
subsequent measurements to the computer,
• After completing a test control following data are displayed in the
workspace:
18.8. Empty – full control
For controls „empty-full" in computer software KTP NET in the product
settings users can set a measurement „charge". The weighing program
during testing requires putting a charge of empty packages and then
a charge of the same packages filled with products put in the same
sequence.
-
Product code and name
-
Product nominal mass
-
Tare value attributed to the product
-
Command concerning putting subsequent products
that are numbered out of the whole sample
-
Number of errors T (Lower than Qn –T) in the sample
-
Number of errors 2T (Lower than Qn –2T) in the sample
-
Sample number
-
Average mass of controlled goods
Test results: OK. - positive, WRONG - negative
-
-
Status
-
Description of button terminating control
-
Net mass of tested products
45