Working Mode - +/- Control; Local Settings; Declaring Checkweighing Thresholds - RADWAG WLC/A1/C/2 User Manual

Precision scale. wlc series.
Table of Contents

Advertisement

• Load the weighing pan with declared amount of parts. When the indication
is stable (
• Single part mass is calculated automatically, next quantity of parts (pcs)
is displayed.
Total weight value of all parts loaded onto the weighing pan
cannot be greater than the max capacity value.
Single part mass value must be equal or greater than 0.1
of the reading unit. Unless this condition is met, the
weighing device displays a message <Err Lo>.
In the course of parts quantity determination before
confirming the declared quantity value it is necessary to wait
for a stable measurement,
22. WORKING MODE – +/- CONTROL
+/- control is a working mode enabling you to enter checkweighing thresholds
values (Min, Max).

22.1. Local Settings

To go to local settings enter <3.3.HiLo> submenu.
Working mode accessibility
3.3.1.Acc
Save mode
3.3.2.Snn
Time interval
3.3.3.Int
Lo threshold
3.3.4.Lo

22.2. Declaring Checkweighing Thresholds

• Enter <HiLo> working mode (+/- control), first, text <SEt Lo> is displayed
for 1 s, next, window for declaring low weighing threshold (Min).
• Enter respective value and press
is displayed for 1 second, next, window for declaring high weighing
threshold (Max).
• Enter respective value and press
home screen is displayed along with declared threshold value, where:
pictogram is displayed), press
key to confirm the mass.
.
For detailed description read section 19.2.1.
For detailed description read section 19.2.2.
For detailed description read section 19.2.3.
For detailed description read section 19.2.4.
key to confirm, first, text <SEt Hi>
key for confirmation, working mode's